Author:
Graham Peter,Howman-Giles Robert,Johnston Ian,Besser Michael
Abstract
✓ Evaluation of 192 cerebrospinal fluid shunts was performed using 1 mCi of technetium-99m DTPA which was injected into the shunt tubing. This was found to be a safe, simple method of evaluating shunt patency. No complications were noted with this procedure, nor was there any documented case of infection related to the injection of the isotope. The sensitivity of the test for evaluation of patency was 97%, specificity 90%, and accuracy 93%. Various patterns of tracer clearance are noted and discussed. The role of this test as an aid in making management decisions is detailed.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Cited by
36 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献