Abstract
Tandem instruments, such as triple-quadrupole or quadrupole-time-of-flight mass spectrometers, often use collisional damping ion guides for the purpose of cooling and focusing the primary ion beam, or as a collision cell for tandem mass spectrometry experiments. A small axial field to reduce the ion residence time in such devices can give considerable improvements in performance. Reduction of the residence time reduces adduct formation caused by unwanted gas-phase reactions and allows multiple reaction monitoring (MRM) and some other types of scan to be carried out more rapidly without the risk of reaction cross-talk. Here we propose a novel electrode arrangement for creation of a suitable axial field in a multipole ion guide without significantly reducing the m/z window. As an example, a second set of four electrodes is added to an existing quadrupole collision cell. The same DC potential is applied to all four extra electrodes, which are shaped in the longitudinal direction to create a suitable axial field inside the device. The potential of these electrodes and the DC bias of the main rods determine the axial field strength. The idea was tested on a modified quadrupole collision cell of a MALDI-QqTOF instrument and experimental results are presented here, as well as a simplified theory of operation.
Subject
Spectroscopy,Atomic and Molecular Physics, and Optics,General Medicine
Cited by
78 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献