Affiliation:
1. The Institute for Cryptosystems, Communication and Informatics, Moscow, Russian Federation
Abstract
We have amplified the model for the drift of ions under a non-uniform, high-frequency electric field by taking space charge effect into account. By this means, we have investigated the effect of space charge on the dynamics of a single type of ions in a spectrometer of ion mobility increment with a cylindrical drift chamber. The counteraction of the space charge effect and the focusing effect is investigated. The output ion current saturation caused by the effect of the space charge is observed. The shape of the ion peak, taking into consideration the space charge effect, has been obtained. We show that the effect of the space charge is sufficient for the relative ion density greater than 10 ppt by order of magnitude (for a cylindrical geometry spectrometer with typical parameters).
Subject
Spectroscopy,Atomic and Molecular Physics, and Optics,General Medicine
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Mass Spectrometry in Russia;European Journal of Mass Spectrometry;2013-12
2. High-Field Asymmetric Waveform Ion Mobility Spectrometry (FAIMS);Practical Aspects of Trapped Ion Mass Spectrometry, Volume IV;2010-05-25
3. Current literature in mass spectrometry;Journal of Mass Spectrometry;2008