Abstract
Background
Heart rate variability (HRV) is used to assess cardiac health and autonomic nervous system capabilities. With the growing popularity of commercially available wearable technologies, the opportunity to unobtrusively measure HRV via photoplethysmography (PPG) is an attractive alternative to electrocardiogram (ECG), which serves as the gold standard. PPG measures blood flow within the vasculature using color intensity. However, PPG does not directly measure HRV; it measures pulse rate variability (PRV). Previous studies comparing consumer-grade PRV with HRV have demonstrated mixed results in short durations of activity under controlled conditions. Further research is required to determine the efficacy of PRV to estimate HRV under free-living conditions.
Objective
This study aims to compare PRV estimates obtained from a consumer-grade PPG sensor with HRV measurements from a portable ECG during unsupervised free-living conditions, including sleep, and examine factors influencing estimation, including measurement conditions and simple editing methods to limit motion artifacts.
Methods
A total of 10 healthy adults were recruited. Data from a Microsoft Band 2 and a Shimmer3 ECG unit were recorded simultaneously using a smartphone. Participants wore the devices for >90 min during typical day-to-day activities and while sleeping. After filtering, ECG data were processed using a combination of discrete wavelet transforms and peak-finding methods to identify R-R intervals. P-P intervals were edited for deletion using methods based on outlier detection and by removing sections affected by motion artifacts. Common HRV metrics were compared, including mean N-N, SD of N-N intervals, percentage of subsequent differences >50 ms (pNN50), root mean square of successive differences, low-frequency power (LF), and high-frequency power. Validity was assessed using root mean square error (RMSE) and Pearson correlation coefficient (R2).
Results
Data sets for 10 days and 9 corresponding nights were acquired. The mean RMSE was 182 ms (SD 48) during the day and 158 ms (SD 67) at night. R2 ranged from 0.00 to 0.66, with 2 of 19 (2 nights) trials considered moderate, 7 of 19 (2 days, 5 nights) fair, and 10 of 19 (8 days, 2 nights) poor. Deleting sections thought to be affected by motion artifacts had a minimal impact on the accuracy of PRV measures. Significant HRV and PRV differences were found for LF during the day and R-R, SDNN, pNN50, and LF at night. For 8 of the 9 matched day and night data sets, R2 values were higher at night (P=.08). P-P intervals were less sensitive to rapid R-R interval changes.
Conclusions
Owing to overall poor concurrent validity and inconsistency among participant data, PRV was found to be a poor surrogate for HRV under free-living conditions. These findings suggest that free-living HRV measurements would benefit from examining alternate sensing methods, such as multiwavelength PPG and wearable ECG.