BACKGROUND
Although computed tomography (CT) studies on machines with more slices have reported higher positive and negative predictive values, the impact of using low-slice (16-slice) CT machines on downstream testing has not been well studied. In community outpatient settings, low-slice CT machines remain in use, although many hospitals have adopted higher-slice machines.
OBJECTIVE
This study examines the association between the use of low-slice CT machines and downstream invasive testing in the context of the CT angiography of the neck.
METHODS
Included health insurance claims pertained to adults with commercial or Medicare Advantage health plans who underwent the CT angiography of the neck. Site certification data were used to assign counts of slices to claims. Claims that were made in the 60 days after CT were examined for cervicocerebral angiography. The association between the number of slices and cervicocerebral angiography was evaluated by using a chi-square test and multivariate logistic regression.
RESULTS
Claims for 16-slice CT had a 5.1% (33/641) downstream cervicocerebral angiography rate, while claims for 64-slice CT had a 3.1% (35/1125) rate, and a significant difference (<i>P</i>=.03) was observed. An analysis that was adjusted for patient demographics also found a significant relationship (odds ratio 1.64, 95% CI 1.00-2.69; <i>P</i>=.047).
CONCLUSIONS
The use of low-slice CT machines in the community may impact the quality of care and result in more downstream testing.