Comparison Between the 24-hour Holter Test and 72-hour Single-Lead Electrocardiogram Monitoring With an Adhesive Patch-Type Device for Atrial Fibrillation Detection: Prospective Cohort Study

Author:

Kwon SoonilORCID,Lee So-RyoungORCID,Choi Eue-KeunORCID,Ahn Hyo-JeongORCID,Song Hee-SeokORCID,Lee Young-ShinORCID,Oh SeilORCID,Lip Gregory Y HORCID

Abstract

Background There is insufficient evidence for the use of single-lead electrocardiogram (ECG) monitoring with an adhesive patch-type device (APD) over an extended period compared to that of the 24-hour Holter test for atrial fibrillation (AF) detection. Objective In this paper, we aimed to compare AF detection by the 24-hour Holter test and 72-hour single-lead ECG monitoring using an APD among patients with AF. Methods This was a prospective, single-center cohort study. A total of 210 patients with AF with clinical indications for the Holter test at cardiology outpatient clinics were enrolled in the study. The study participants were equipped with both the Holter device and APD for the first 24 hours. Subsequently, only the APD continued ECG monitoring for an additional 48 hours. AF detection during the first 24 hours was compared between the two devices. The diagnostic benefits of extended monitoring using the APD were evaluated. Results A total of 200 patients (mean age 60 years; n=141, 70.5% male; and n=59, 29.5% female) completed 72-hour ECG monitoring with the APD. During the first 24 hours, both monitoring methods detected AF in the same 40/200 (20%) patients (including 20 patients each with paroxysmal and persistent AF). Compared to the 24-hour Holter test, the APD increased the AF detection rate by 1.5-fold (58/200; 29%) and 1.6-fold (64/200; 32%) with 48- and 72-hour monitoring, respectively. With the APD, the number of newly discovered patients with paroxysmal AF was 20/44 (45.5%), 18/44 (40.9%), and 6/44 (13.6%) at 24-, 48-, and 72-hour monitoring, respectively. Compared with 24-hour Holter monitoring, 72-hour monitoring with the APD increased the detection rate of paroxysmal AF by 2.2-fold (44/20). Conclusions Compared to the 24-hour Holter test, AF detection could be improved with 72-hour single-lead ECG monitoring with the APD.

Publisher

JMIR Publications Inc.

Subject

Health Informatics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3