Abstract
The accurate and precise HPLC analytical method validated for the determination of Amlodipine besylate in pharmaceutical dosage form.The chromatographic separation is carried out on shimadzu HPLC system (LC-2010 CHT) with UV Vissible detector and C18(150mm x3.9 mm) 5 μm Column. The Mobile phase consists of Acetonitrile: Methanol: PH 3.0 Buffer (15 V: 35 V: 50 V) , at the flow rate of 1.0 ml/min and elutes were monitoring at 237 nm. The observed retention time for Amlodipine besylate was 10.8 min. The % RSD for system precision was 0.41 % and Method precision was 0.58 %. The method was found to linear (R=0.99996) in the Concentration range of 35-105 μg/ml (50 to 150%). The accuracy was in between 99.50-99.91%.
Keywords: HPLC, Correlation coefficient, System suitability, Bias, % RSD and ICH guidelines