Author:
D PATEL BHOOMI,BHAVYA MEHTA,B CHAUDHARY ANKIT
Abstract
Objective: The objective of the study was to develop and validate reverse-phase high-performance liquid chromatography (RP-HPLC) method and apply method to tablet dosage form.
Methods: A simple, rapid, economical, precise, and accurate RP-HPLC method for simultaneous estimation of lamivudine and zidovudine in their combined dosage form has been developed.
Results: A RP-HPLC method was developed for the simultaneous estimation of lamivudine and zidovudine. In their combined dosage form has been developed. The separation was achieved by LC-C18 column (150 mm ×4.6 mm, 5 μm) and water: methanol (65:35v/v) as mobile phase, at a flow rate of 0.8 ml/min. Detection was carried out at 272 nm. Retention time of lamivudine and zidovudine was found to be 3.007 min and 4.647, respectively. The method has been validated for linearity, accuracy, and precision. The assay method was found to be linear from 50% to 150% for lamivudine and zidovudine.
Conclusion: Developed method was found to be accurate, precise, and rapid for simultaneous estimation of lamivudine and zidovudine in their combined dosage form.
Publisher
Innovare Academic Sciences Pvt Ltd
Subject
Pharmacology (medical),Pharmaceutical Science,Pharmacology
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献