Abstract
Objectives: The objective of the present work was to develop an analytical method and validation for the estimation of Ezetimibe in rabbit plasma using high-performance liquid chromatography (HPLC).
Methods: A simple, rapid, sensitive, and accurate HPLC method was developed and validated for the quantification of Ezetimibe concentration in rabbit plasma using metoclopramide as an internal standard. Separation was performed on the Xerra C18 column (250 × 4.6 mm 5 μm) using a mobile phase consisting of 0.1% perchloric acid: acetonitrile(55:45 v/v) at a flow rate of 1 ml/min. Validation of the method was performed to demonstrate its selectivity, linearity, precision, accuracy, ruggedness, recovery, and matrix effect.
Results: The calibration curves of Ezetimibe were linear over a concentration range of 5–1022 μg/mL. The with-in and between-day coefficients of variation were <10%. The extraction recoveries of Ezetimibe at the three levels of quality control samples were 99.961%, 99.767%, and 99.938%.
Conclusion: The method was rapid with a retention time of Ezetimibe and the internal standard observed at 10 min, respectively. The developed method was successfully applied to studying the pharmacokinetics of Ezetimibe in rabbits.
Publisher
Innovare Academic Sciences Pvt Ltd
Reference6 articles.
1. Hammady T, El-Gindy A, Lejmi E, Dhanikula RS, Moreau P, Hildgen P. Characteristics and properties of nanospheres co-loaded with lipophilic and hydrophilic drug models. Int J Pharm. 2009;369: 185-95.
2. Elbary AA, Ali AA, Aboud HM. Enhanced dissolution of meloxicam from orodispersible tablets prepared by different methods. Bull Fac Pharm Cario Univ. 2012;50:89-97.
3. Rani TN, Muzib YI. Rapid melts: A review. Int J Pharm Chem Sci. 2014;3:118-30.
4. Rani TN, Muzib YI. Formulation and in vitro evaluation of ezetimibe rapidmelts. Asian J Pharm Clin Res. 2020;13:97-103.
5. Mohanachandran PS, Krishna Mohan PR, Saju F, Bini KB, Babu B, Shalina KK. Formulation and evaluation of mouth dispersible tablets of amlodipine besylate. Int J Appl Pharm 2010;2(3):1-6.