OCULAR DRUG DELIVERY SYSTEM USING OPEN-SOURCE SYRINGE PUMP

Author:

V Surya,Bose Srutartha,Kulkarni Shreya,D Kathirvelu

Abstract

Objective: The objective of this study is to design an automated ocular drug delivery system controlled by an open-source syringe pump, which in turn is to be controlled using Arduino UNO.Methods: The Arduino UNO is the microcontroller used to control the entire setup. The drug delivery is executed using the open-source syringe pump. A 12V NEMA 17 stepper motor is used to execute the pump movements. To facilitate the ease of use, an Android application developed using Android Studio has been developed, where inputs are given in the form of the required drug dose. The application plays the role of converting the requested drug dose into the number of steps required to run the stepper motor. The stepper motor works based on simple lead-screw mechanism.Results: The designed syringe pump has been tested using water. It was determined that for 42 steps a drug volume of 30 μl is dispensed by the designed device. Typically, a drop volume ranges from 33.8 μl to 63.4 μl and the device that has been discussed in this article can deliver a minimum volume of 30 μl.Conclusion: Using this device, excessive use of drugs can be minimalized, thus reducing the capital investment put behind ocular drugs.

Publisher

Innovare Academic Sciences Pvt Ltd

Subject

Pharmacology (medical),Pharmaceutical Science,Pharmacology

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3