Alanine Aminotransferase Results Differ by Analyzer Manufacturer in a National Integrated Health Setting, 2012–2017

Author:

Beste Lauren A.1,Icardi Michael1,Hunt Christine M.1,Gylys-Colwell Ina1,Lowy Elliott1,Taylor Leslie1,Morgan Timothy R.1,Chang Michael F.1,Maier Marissa M.1,Cheung Ramsey1

Affiliation:

1. From the General Medicine Service (Dr Beste) and Health Services Research and Development (Drs Beste, Lowy, and Taylor, and Ms Gylys-Colwell), VA Puget Sound Health Care System, Seattle, Washington; General Internal Medicine, University of Washington School of Medicine, Seattle (Dr Beste); Pathology Service, VA Iowa City, Iowa City (Dr Icardi); Pathology Department, University of Iowa Carver Coll

Abstract

Context.— Disease guidelines specify universal alanine aminotransferase (ALT) thresholds for clinical decision-making, yet the effect of variability among ALT analyzers remains unclear. Objective.— To compare ALT results from different analyzers from 2012–2017. Design.— Veterans Health Administration (VHA) laboratories perform external ALT proficiency testing using standardized College of American Pathologists (CAP) samples in analyzers by 5 manufacturers. In this operational analysis, we evaluated 22 950 ALT values from 80 independent CAP samples tested at 223 laboratories. Using mixed effects modeling, we estimated the association between analyzer manufacturer and CAP outcome, adjusting for manufacturer, facility, and calendar year. We performed subgroup analyses on CAP samples with overall means near clinical guideline-specified thresholds, including less than 50 U/L (n = 10) and less than 35 U/L (n = 5). Results.— The VHA used Abbott Laboratories (n = 3175; 14%), Beckman Coulter Diagnostics (n = 8723; 38%), Roche Diagnostics (n = 2595; 11%), Siemens Healthineers USA (n = 5713; 25%), and Vitros/Ortho Clinical Diagnostics (n = 2744; 12%) analyzers. The CAP samples (n = 80 samples, n = 22 950 tests) covered a wide range of mean ALT values (21–268 U/L). The average difference in mean ALT value per sample between the highest-reading and lowest-reading manufacturers was 15.4 U/L (SD = 1.8) for the 10 samples with mean ALT less than 50 U/L, and it was 10.4 U/L (SD = 3.6) overall (n = 80). In linear mixed effects modeling, we found statistically significant differences in ALT values between the different manufacturers in each year. Conclusions.— We found statistically and clinically meaningful differences between analyzers across the ALT spectrum in each year, including at ALT levels lower than 50 U/L and lower than 35 U/L. Universal ALT thresholds should be avoided as a trigger for clinical action until differences between analyzers can be resolved.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3