Traceability Assessment and Performance Evaluation of Results for Measurement of Abbott Clinical Chemistry Assays on 4 Chemistry Analyzers

Author:

Lim Jinsook,Song Kyung Eun,Song Sang Hoon,Choi Hyun-Jung,Koo Sun Hoe,Kwon Gye Choel1

Affiliation:

1. From the Department of Laboratory Medicine, Chungnam National University Hospital, Daejeon, South Korea (Drs Lim, Koo, and Kwon); the Department of Laboratory Medicine, Kyungpook National University Hospital, Daegu, South Korea (Dr K.E. Song); the Department of Laboratory Medicine, Seoul National University Hospital and College of Medicine, Seoul, South Korea (Dr S.H. Song); the Department of Lab

Abstract

Context.—The traceability of clinical results to internationally recognized and accepted reference materials and reference measurement procedures has become increasingly important. Therefore, the establishment of traceability has become a mandatory requirement for all in vitro diagnostics devices. Objectives.—To evaluate the traceability of the Abbott Architect c8000 system (Abbott Laboratories, Abbott Park, Illinois), consisting of calibrators and reagents, across 4 different chemistry analyzers, and to evaluate its general performance on the Toshiba 2000FR NEO (Toshiba Medical Systems Corporation, Otawara-shi, Tochigi-ken, Japan). Design.—For assessment of traceability, secondary reference materials were evaluated 5 times, and then bias was calculated. Precision, linearity, and carryover were determined according to the guidelines of the Clinical and Laboratory Standards Institute (Wayne, Pennsylvania). Results.—The biases from 4 different analyzers ranged from −2.33% to 2.70% on the Toshiba 2000FR NEO, −2.33% to 5.12% on the Roche Hitachi 7600 (Roche Diagnostics International, Basel, Switzerland), −0.93% to 2.87% on the Roche Modular, and −2.16% to 2.86% on the Abbott Architect c16000. The total coefficients of variance of all analytes were less than 5%. The coefficients of determination (R2) were more than 0.9900. The carryover rate ranged from −0.54% to 0.17%. Conclusions.—Abbott clinical chemistry assays met the performance criteria based on desirable biological variation for precision, bias, and total error. They also showed excellent linearity and carryover. Therefore, these clinical chemistry assays were found to be accurate and reliable and are readily applicable on the various platforms used in this study.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3