Performance Evaluation of the Sysmex XE-5000 Hematology Analyzer for White Blood Cell Analysis in Cerebrospinal Fluid

Author:

Perné Andrea,Hainfellner Johannes A.,Womastek Irene,Haushofer Alexander,Szekeres Thomas,Schwarzinger Ilse

Abstract

Context.—The newest generation hematology analyzer, Sysmex XE-5000 (Sysmex Corporation, Kobe, Japan) is equipped with an improved body fluid analysis mode. Objective.—To evaluate the applicability of the XE-5000 analyzer to white blood cell (WBC) analysis in cerebrospinal fluid (CSF). Design.—A total of 425 routinely collected, consecutive CSF samples were included in the study. For a comparison of total WBC counts, the results of routine chamber counts were grouped into categories of 0 to 5 (n  =  330), >5 to 10 (n  =  36), >10 to 50 (n  =  39), >50 to 200 (n  =  15), and >200 (n  =  5) WBC/µL. Microscopic differential counts were performed using cytospins from 276 samples. Results were grouped according to the percent content of polymorphonuclear (PMN) cells, 0% to 25% (n  =  263), >25% to 50% (n  =  7), >50% to 75% (n  =  3), and >75% to 100% (n  =  3) of WBC. Corresponding results of XE-5000 analysis were matched to these particular count categories. Results.—For total WBC counts, the proportions of samples correctly classified by the XE-5000 from the percentage groups described above were 88%, 47%, 72%, 93%, and 100%, respectively. After the two lowest count categories were combined into one range of 0 to 10 WBC/µL, matches increased to 95%. For PMN counts in the 0% to 25% group, 37% of samples were misclassified by the XE-5000. Conversely, for samples with microscopic PMN counts of more than 25%, there was a trend toward underestimation by the XE-5000. Mismatches were most pronounced in samples with fewer than 10 WBC/µL. Conclusions.—The Sysmex XE-5000 hematology analyzer yields valid total CSF cell counts and may be considered an acceptable alternative to the traditional chamber method, even for samples with low WBC counts. However, it cannot be recommended as a suitable alternative for manual differential cytologic workup.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

Cited by 29 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3