Validation and Optimization of Criteria for Manual Smear Review Following Automated Blood Cell Analysis in a Large University Hospital

Author:

Pratumvinit Busadee1,Wongkrajang Preechaya1,Reesukumal Kanit1,Klinbua Cherdsak1,Niamjoy Patama1

Affiliation:

1. From the Department of Clinical Pathology, Faculty of Medicine, Siriraj Hospital, Mahidol University, Bangkok, Thailand.

Abstract

Context.—Each laboratory should have criteria for manual smear review that limit workload without affecting patient care. The International Consensus Group for Hematology Review established guidelines for action after automated blood cell analysis in 2005. Objective.—To compare the consensus group criteria with our laboratory criteria and optimize them for better efficiency. Design.—A total of 2114 first-time samples were collected consecutively from daily workload and were used to compare 2 criteria as well as establish the optimized criteria. Another set of 891 samples was used to validate the optimized criteria. All samples were run on either Sysmex XE-5000 or Coulter LH750 hematology analyzers and were investigated by manual smear review. The efficiency of each set of criteria was compared and optimized to obtain better efficiency, an acceptable review rate, and a low false-negative rate. Results.—From 2114 samples, 368 (17.40%) had positive smear results. Compared with that of our laboratory criteria, the efficiency of the consensus group criteria was higher (83.63% versus 78.86%, P < .001), the review rate was higher (29.33% versus 22.37%, P < .001), and the false-negative rate was lower (2.22% versus 8.09%, P < .001). After optimizing the rules, we obtained an efficiency of 87.13%, a review rate of 24.22%, and a false-negative rate of 2.98%. We validated the optimized criteria with another set of samples, and the efficiency, review rate, and false-negative rate were 87.32%, 25.25%, and 1.12%, respectively. Conclusions.—Each laboratory should verify the criteria for smear review, based on the International Consensus Group for Hematology Review, and optimize them to maximize efficiency.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

Cited by 29 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3