Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature

Author:

Gulati Gene L.1,Hyland Lawrence J.1,Kocher William1,Schwarting Rolland1

Affiliation:

1. From the Department of Pathology, Anatomy, and Cell Biology, Jefferson Medical College, Thomas Jefferson University, Philadelphia, Pa

Abstract

Abstract Objective.—To delineate changes that occur in various parameters of automated complete blood cell count (CBC) and differential leukocyte count (differential) on prolonged storage of blood at room temperature. Design.—A CBC and an automated differential were performed on the Coulter Gen.S on 40 K3 (tripotassium ethylenediamine-tetraacetate) EDTA-anticoagulated blood specimens once daily, specimen volume permitting, for 3 to 7 days. Specimens were kept at room temperature throughout the study. The results were tabulated using a personal computer with Excel software. Percent change or absolute difference from the initial value for each parameter for each subsequent day of the study period was calculated. Results.—Among the CBC parameters, hemoglobin, red blood cell count, and mean corpuscular hemoglobin were stable for the duration of the study (7 days), white blood cell count was stable for at least 3 days (up to 7 days, if the count was within or above the normal range), and platelet count was stable for at least 4 days (up to 7 days, if the count was within or above the normal range). The mean corpuscular volume, mean platelet volume, hematocrit, and red blood cell distribution width each increased, and the mean corpuscular hemoglobin concentration decreased from day 2 onward. Among the differential parameters, the relative percentages and absolute numbers of neutrophils, lymphocytes, and eosinophils tended to increase, whereas those of monocytes trended downward over time. Limited data on basophils did not reveal an appreciable change. Conclusion.—Blood specimens stored at room temperature for more than 1 day (up to 3 days or possibly longer) were found to be acceptable with some limitations for CBC but not for the differential.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3