Micro–Red Blood Cell, Fragmented Red Blood Cell, Platelet Distribution Width, Mean Platelet Volume, and Platelet–Large Cell Ratio on Sysmex XN Series Hematology Analyzers Can Be Used for the Reflex Test of Impedance Platelet Count in Clinical Practice

Author:

Chen Si1,Mao Zhigang1,Wang Shuang1,Deng Jiamin2,Liao Hongyan1,Zheng Qin1

Affiliation:

1. From the Department of Laboratory Medicine, West China Hospital of Sichuan University, Chengdu, China (Chen, Mao, Wang, Liao, Zheng)

2. West China School of Medicine, Sichuan University, Chengdu, China (Deng)

Abstract

Context.— Platelet (PLT) counting with impedance (PLT-I) is widely used but has low specificity. PLT counting with fluorescence (PLT-F), tested by the Sysmex XN series with high specificity, can be a complementary method to PLT-I. Objective.— To identify red blood cell (RBC)– and PLT-related parameters as potential influencing factors for PLT-I and establish PLT reflex test rules with PLT-F. Design.— We prospectively tested both PLT-I and PLT-F in all 3480 samples. In a development data set of 3000 samples, differences between the reflex and nonreflex groups were compared and influencing factors for PLT-I were identified by logistic regression. The area under the receiver operating characteristic (ROC) curve and cutoff values were obtained by ROC curve analysis. Validation was conducted in the remaining 480 samples (validation data set). Results.— PLT-F showed comparable results with immunoplatelet counting. In logistic regression, increased micro-RBC absolute count (micro-RBC#), fragmented RBC absolute count (FRC#), PLT distribution width (PDW), mean PLT volume (MPV), PLT–large cell ratio (P-LCR), and immature PLT fraction absolute count (IPF#) were influencing factors for PLT-I. In ROC curve analysis, the cutoff values of micro-RBC#, FRC#, PDW, MPV, and P-LCR were 0.64 × 106/μL, 0.082 × 106/μL, 15.40 fL, 11.15 fL, and 33.95%, respectively. The areas under the ROC curve of micro-RBC# and FRC# were 0.77 and 0.79, respectively. Conclusions.— Micro-RBC#, FRC#, PDW, MPV, P-LCR, and IPF# were factors affecting PLT-I. Among them, micro-RBC# and FRC# were the most impactful factors. From our study results, micro-RBC#, FRC#, MPV, PDW, and P-LCR can be used to establish reflex test rules for PLT counting in clinical work.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3