Estimated Risk for Insulin Dose Error Among Hospital Patients Due to Glucose Meter Hematocrit Bias in 2020

Author:

Inman Mark1,Lyon Andrew W.2,Lyon Oliver A. S.3,Lyon Martha E.2

Affiliation:

1. From the Department of Pediatrics, University of Saskatchewan, Saskatoon, Saskatchewan, Canada (Inman)

2. The Department of Pathology & Laboratory Medicine, Division of Clinical Biochemistry, Saskatchewan Health Authority, Saskatoon, Saskatchewan, Canada (AW Lyon, ME Lyon)

3. School of Computing, Queen's University, Kingston, Ontario, Canada (OAS Lyon)

Abstract

Context.—Glycemic control requires accurate blood glucose testing. The extent of hematocrit interference is difficult to assess to assure quality patient care.Objective.—To predict the effect of patient hematocrit on the performance of a glucose meter and its corresponding impact on insulin-dosing error.Design.—Multilevel mixed regression was conducted to assess the extent that patient hematocrit influences Roche Accu-Chek Inform II glucose meters, using the Radiometer ABL 837 as a reference method collected during validation of 35 new meters. Regression coefficients of fixed effects for reference glucose, hematocrit, an interaction term, and random error were applied to 4 months of patient reference method results extracted from the laboratory information system. A hospital inpatient insulin dose algorithm was used to determine the frequency of insulin dose error between reference glucose and meter glucose results.Results.—Fixed effects regression for method and hematocrit predicted biases to glucose meter results that met the “95% within ±12%” for the US Food and Drug Administration goal, but combinations of fixed and random effects exceeded that target in emergency and hospital inpatient units. Insulin dose errors were predicted from the meter results. Twenty-eight percent of intensive care unit, 20.8% of hospital inpatient, and 17.7% of emergency department results were predicted to trigger a ±1 insulin dose error by fixed and random effects.Conclusions.—The current extent of hematocrit interference on glucose meter performance is anticipated to cause insulin error by 1-dose category, which is likely associated with low patient risk.

Publisher

Archives of Pathology and Laboratory Medicine

Subject

Medical Laboratory Technology,General Medicine,Pathology and Forensic Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3