Abstract
Abstract
In this study, a dual-band hybrid filter that achieves high excitation light rejection performance in a lensless imaging system was fabricated and incorporated into an imaging device. The hybrid filter consisted of interference and absorption filters, and a fiber optic plate (FOP). The interference filters were attached to both sides of the FOP, which was placed on top of the absorption filter to suppress the decrease in spatial resolution. In addition, the lamination order was optimized to achieve a high fluorescence observation performance. The fabricated hybrid filter was mounted on an image sensor and had the ability to indicate the green and red fluorescence components.
Subject
General Physics and Astronomy,Physics and Astronomy (miscellaneous),General Engineering
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献