Author:
Shoji Yuto,Mori Shohei,Arakawa Mototaka,Ohba Shigeo,Kobayashi Kazuto,Kanai Hiroshi
Abstract
Abstract
We developed a single ultrasound probe to simultaneously measure blood pressure and changes in the diameter of the radial artery to estimate the wall viscoelasticity during flow-mediated dilatation (FMD). This probe can be used for the early diagnosis of arteriosclerosis. This paper introduces the pulse transit time method to accurately measure changes in blood pressure during FMD. Using the single ultrasound probe and the proposed method, in vivo experiments involving three subjects were conducted, and reasonable results on blood pressure were obtained. Thus, the usefulness of the pulse transit time method was experimentally confirmed.
Subject
General Physics and Astronomy,Physics and Astronomy (miscellaneous),General Engineering
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献