Author:
Sun Hui,Wang Wei,Zhang Zixin,Wang Li,Zhang Wuyu,Xiong Kedi,Yang Sihua
Abstract
Abstract
Currently, the reported endoscopic system cannot meet the necessary conditions for real-time and optical-resolution clinical application simultaneously. In this study, by utilizing a high-repetition-rate laser and optimizing the overall structure of the probe, a real-time optical-resolution photoacoustic endoscope was developed, which could image targets in real time while maintaining a relatively constant lateral resolution. The endoscopic system has a best-resolution of 19 μm and a real-time imaging speed of 25 Hz. Experimental results demonstrate that this approach will push the promotion and commercialization of photoacoustic endoscopic imaging in a variety of preclinical and clinical applications.
Funder
National Natural Science Foundation of China
The Science and Technology Planning Project of Guangdong Province, China
Science and Technology Program of Guangdong
Postdoctoral Science Foundation of China
Natural Science Foundation of Guangdong Province
Subject
General Physics and Astronomy,General Engineering
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献