Identification of Factors That Disrupt Negative Air Pressurization of Respiratory Isolation Rooms

Author:

Pavelchak Nicholas,DePersis Ronald P.,London Matthew,Stricof Rachel,Oxtoby Margaret,DiFerdinando George,Marshall Elizabeth

Abstract

AbstractObjectives:To investigate the airflow characteristics of respiratory isolation rooms (IRs) and to evaluate the use of visible smoke as a monitoring tool.Methods:Industrial hygienists from the New York State Department of Health evaluated 140 designated IRs in 38 facilities within New York State during 1992 to 1998. The rooms were located in the following settings: hospitals (59%), correctional facilities (40%), and nursing homes (1%). Each room was tested with visible smoke for directional airflow into the patient room (ie, negative air pressure relative to adjacent areas). Information was obtained on each facility's policies and procedures for maintaining and monitoring the operation of the IRs.Results:Inappropriate outward airflow was observed in 38% of the IRs tested. Multiple factors were associated with outward airflow direction, including ventilation systems not balanced (54% of failed rooms), shared anterooms (14%), turbulent airflow patterns (11%), and automated control system inaccuracies (10%). Of the 140 tested rooms, 38 (27%) had either electrical or mechanical devices to monitor air pressurization continuously. The direction of airflow at the door to 50% (19/38) of these rooms was the opposite of that indicated by the continuous monitors at the time of our evaluations. The inability of continuous monitors to indicate the direction of airflow was associated with instrument limitations (74%) and malfunction of the devices (26%). In one facility, daily smoke testing by infection control staff was responsible for identifying the malfunction of a state-of-the-art computerized ventilation monitoring and control system in a room housing a patient infectious with drug-resistant tuberculosis.Conclusion:A substantial percentage of IRs did not meet the negative air pressure criterion. These failures were associated with a variety of characteristics in the design and operation of the IRs. Our findings indicate that a balanced ventilation system does not guarantee inward airflow direction. Devices that continuously monitor and, in some cases, control the pressurization of IRs had poor reliability. This study demonstrates the utility of using visible smoke for testing directional airflow of IRs, whether or not continuous monitors are used. Institutional tuberculosis control programs should include provisions for appropriate monitoring and maintenance of IR systems on a frequent basis, including the use of visible smoke.

Publisher

Cambridge University Press (CUP)

Subject

Infectious Diseases,Microbiology (medical),Epidemiology

Reference11 articles.

1. New York State Public Health Law, Title 10 NYCRR, section 712.31, Table 3 (February 29, 1988).

2. Notice of proposed rule-making on Occupational Exposure to Tuberculosis (29 CFR 1910);Federal Register,1997

Cited by 37 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3