Author:
Gangneux Jean-Pierre,Robert-Gangneux Florence,Gicquel Guirec,Tanquerel Jean-Jacques,Chevrier Sylviane,Poisson Magali,Aupée Martine,Guiguen Claude
Abstract
We compared the yields of 4 recently developed sieve impactor air samplers that meet international standard ISO 14698-1, using 2 growth media (tryptic soy agar and malt extract agar) in real conditions of use. Several hospital sites expected to have different densities of airborne microflora were selected in 2 hospitals. The Samplair MK2, Air Ideal, and Mas-100 samplers yielded higher bacterial counts than did the SAS Super-100 device (P<.05). No significant differences in fungal counts were noted between the 4 devices. The use of malt extract agar in addition to tryptic soy agar significantly improved the fungal yield.
Publisher
Cambridge University Press (CUP)
Subject
Infectious Diseases,Microbiology (medical),Epidemiology
Cited by
28 articles.
订阅此论文施引文献