Adequate Disinfection of a Split-Septum Needleless Intravascular Connector with a 5-Second Alcohol Scrub

Author:

Rupp Mark E.,Yu Stephanie,Huerta Tomas,Cavalieri R. Jennifer,Alter Roxanne,Fey Paul D.,Van Schooneveld Trevor,Anderson James R.

Abstract

Objective.Define optimum vascular catheter connector valve disinfection practices under laboratory and clinical conditions.Design.Prospective observational clinical survey and laboratory assessment of disinfection procedures.Setting.All adult inpatients at an academic healthcare center.Methods.In the clinical setting, contamination of needleless connectors was assessed in 6 weekly prevalence surveys in which the connector valves from central venous catheters (CVCs) in situ were cultured by pressing the connector diaphragm to an agar plate. Before culture, valves were disinfected by scrubbing the diaphragm with a 70% isopropyl alcohol pledget for 0, 5, 10, 15, or 30 seconds. In the laboratory, the diaphragms on 150 unused sterile connector valves were inoculated with 103, 105, or 108colony-forming units ofStaphylococcus epidermidisand allowed to dry. After disinfection of the diaphragms by scrubbing with a 70% isopropyl alcohol pledget for 0, 5, 10, 15, or 30 seconds, the valves were sampled by pressing the diaphragm to an agar plate.Results.In the clinical setting, 363 connector valves from patients with CVCs were sampled, and 66.7% of nondisinfected valves revealed bacterial contamination. After 5-second disinfection with an alcohol pledget, only 1 (1.4%) of 71 yielded microbial growth (P< .005). In the laboratory, at the 103and 105inoculum, all connector valves yielded sterile cultures when scrubbed for 5 or more seconds (P< .001). At the 108inoculum, 2 (20%) of 10 connector valves yielded minimal growth ofS. epidermidis.Conclusions.A 5-second scrub with a 70% isopropyl alcohol pledget yields adequate disinfection of a split-septum intravascular catheter connector valve under clinical and laboratory conditions.

Publisher

Cambridge University Press (CUP)

Subject

Infectious Diseases,Microbiology (medical),Epidemiology

Cited by 38 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3