Application of Environmental Sampling to Flexible Endoscope Reprocessing: The Importance of Monitoring the Rinse Water

Author:

Muscarella Lawrence F.

Abstract

AbstractThe routine sampling of environmental surfaces within a healthcare facility is generally not recommended by the Centers for Disease Control and Prevention (CDC), the Association for the Advancement of Medical Instrumentation (AAMI), and several other healthcare organizations. There are a few circumstances, however, for which some organizations do recommend this practice. For instance, the CDC and the Association for Professionals in Infection Control and Epidemiology (APIC) recommend environmental sampling as clinically required during an outbreak investigation. The CDC and AAMI also recommend routine sampling of the rinse water used during hemodialyzer (but not endoscope) reprocessing. The rationale for this recommendation is based in part on reports of pyrogenic responses, patient infections, and bacteremia due to waterborne, gram-negative bacteria during hemodialysis. To determine whether the basis for this rationale might similarly apply to the rinse water used during endoscope reprocessing, the Food and Drug Administration's medical device reporting database, the endoscope reprocessing literature, and other sources were reviewed. The results of this review indicate that nosocomial outbreaks linked to endoscopes contaminated with gram-negative bacteria have been frequently reported. As a result, for several reasons, including to minimize the risk of patient infection due to gram-negative bacteria following endoscopy, this article recommends routine microbiologic sampling of the rinse water used during endoscope reprocessing.

Publisher

Cambridge University Press (CUP)

Subject

Infectious Diseases,Microbiology (medical),Epidemiology

Cited by 38 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. ANSI/AAMI ST108:2023; Water for the processing of medical devices;ANSI/AAMI ST108:2023; Water for the processing of medical devices;2023-06-30

2. Final rinse water quality for flexible endoscopy to minimize the risk of post-endoscopic infection. Report from Healthcare Infection Society Working Party;Journal of Hospital Infection;2022-06

3. Further Reading;Sterile Processing of Pharmaceutical Products;2022-01-07

4. SAFETY IN THE GASTROENTEROLOGY SETTING;Gastroenterology Nursing;2021-09-09

5. Infection Control and Radiation Safety in the Bronchoscopy Suite;Flexible Bronchoscopy;2020-05-08