Impact of heterozygous hemoglobin E on six commercial methods for hemoglobin A1c measurement

Author:

Yong Sharon1,Liu Hong1,Lum Cindy Lye Teng2,Liu Qian2,Sim Sin Ye3ORCID,Chay Felicia Fu Mun3,Cheng Wan Ling4,Neo Siew Fong4,Chew Suru4,Ong Lizhen4,Loh Tze Ping4,Liu Qinde1,Teo Tang Lin1,Sethi Sunil Kumar4

Affiliation:

1. Chemical Metrology Division, Health Sciences Authority, Singapore, Singapore

2. Department of Pathology, Sengkang General Hospital, Singapore, Singapore

3. Department of Laboratory Medicine, Alexandra Hospital, Singapore, Singapore

4. Department of Laboratory Medicine, National University Hospital, Singapore, Singapore

Abstract

Background This study examined the impact of heterozygous HbE on HbA1c measurements by six commonly used commercial methods. The results were compared with those from a modified isotope-dilution mass spectrometry (IDMS) reference laboratory method on a liquid chromatograph coupled with tandem mass spectrometer (LC-MS/MS). Methods Twenty-three leftover samples of patients with heterozygous HbE (HbA1c range: 5.4–11.6%), and nineteen samples with normal hemoglobin (HbA1c range: 5.0–13.7%) were included. The selected commercial methods included the Tina-quant HbA1c Gen. 3 (Roche Diagnostics, Basel, Switzerland), Cobas B 101 (Roche Diagnostics, Basel, Switzerland), D100 (Bio-Rad Laboratories, Hercules, CA, USA), Variant II Turbo HbA1c 2.0 (Bio-Rad Laboratories, Hercules, CA, USA), DCA Vantage (Siemens Healthcare, Erlangen, Germany) and HbA1c Advanced (Beckman Coulter Inc., Brea, CA, USA). Results With the exception of Cobas B 101 and the Variant II Turbo 2.0, the 95% confidence intervals of the Passing–Bablok regression lines between the results from the six commercial methods and the IDMS method overlapped. The latter suggested no statistically significant difference in results and hence no impact on HbA1c result despite the presence of heterozygous HbE. The method of Cobas B 101 gave positive bias at the range of concentrations examined (5.4–11.6%), while that of Variant II Turbo 2.0 gave positive bias at concentrations up to approximately 9.5%. The finding of significant positive bias in the methods of Cobas B 101 and Variant II Turbo 2.0 agrees with the observations of some previous studies, but is contrary to manufacturer’s claim indicating the absence of interference by heterozygous HbE. Our results also clearly showed the impact of heterozygous HbE across a fairly broad measurement range using a laboratory method (the Variant II Turbo 2.0). Laboratory practitioners and clinicians should familiarize themselves with prevailing hemoglobin variants in the population they serve and select the appropriate methods for HbA1c measurement.

Publisher

PeerJ

Reference13 articles.

1. Glycemic targets: standards of medical care in diabetes-2019;American Diabetes Association;Diabetes Care,2009

2. Hemoglobin E. Orphanet encyclopedia;Bachir,2004

3. Impact of analytical and biological variations on classification of diabetes using fasting plasma glucose, oral glucose tolerance test and HbA1c;Chai;Scientific Reports,2017

4. College of American Pathologists (CAP) GH5 survey data;College of American Pathologists,2020

5. 510(k) substantial equivalence determination decision summary for cobas b 101 system;Food and Drug Administration,2020