Efficacy of extraoral suction devices in aerosol and splatter reduction during ultrasonic scaling: A laboratory investigation

Author:

Horsophonphong Sivaporn1ORCID,Chestsuttayangkul Yada2,Surarit Rudee3ORCID,Lertsooksawat Wannee4ORCID

Affiliation:

1. Department of Pediatric Dentistry, Faculty of Dentistry, Mahidol University, Bangkok Thailand

2. Dental Department, Rajavithi Hospital, Bangkok Thailand

3. Department of Oral Biology, Faculty of Dentistry, Mahidol University, Bangkok Thailand

4. Department of Pharmacology, Faculty of Dentistry, Mahidol University, Bangkok Thailand

Abstract

Background. Ultrasonic scaling generates aerosols and splatters contaminated with microorganisms, increasing the risk of disease transmission in the dental office. The present study aimed to evaluate the effectiveness of extraoral suction (EOS) units in aerosol and splatter reduction during ultrasonic scaling. Methods. Ultrasonic scaling was conducted on a dental manikin headset to simulate a scaling procedure. Water containing Lactobacillus acidophilus at a concentration of 107 colony-forming units per milliliter and 1% fluorescein solution was used as the water supply of the scaler. The scaling procedure was conducted with a high-volume evacuator (HVE) or the combination of HVE and an EOS unit. de Man–Rogosa–Sharpe agar plates were placed at different distances surrounding the dental chair. Filter papers were placed at various positions surrounding the oral cavity and on areas of the body. Results. Bioaerosols were detected at every sampling site and could travel as far as 150 cm from the oral cavity. The combination of HVE and EOS significantly reduced the total number of bacterial colonies in the air (P<0.001). Dissemination of the stain was in the range of 20 cm from the oral cavity. The maximum contaminated surface area was at the 4 o’clock position from the oral cavity. The combination of EOS and HVE significantly reduced the contaminated area (P<0.05). The stain was also found on the wrists, chest, abdomen, and lap of the operator and assistant. The lap was the most contaminated area of the body. Conclusion. EOS was effective in reducing the bioaerosols and splatters generated during ultrasonic scaling.

Publisher

Maad Rayan Publishing Company

Subject

General Dentistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3