When the blood-back detection system fails: an IABP case report

Author:

Linley G H1,Bakker E W1,de Vroege R2,Spijkstra J J3

Affiliation:

1. Department of Extracorporeal Circulation, Vrije Universiteit Medisch Centrum, Amsterdam, The Netherlands

2. Department of Extracorporeal Circulation, Vrije Universiteit Medisch Centrum, Amsterdam, The Netherlands,

3. Department of Intensive Care, Vrije Universiteit Medisch Centrum, Amsterdam, The Netherlands

Abstract

A 74-year-old male patient’s circulation was supported by an intra-aortic balloon for a period of six days following multiple cardiac infarcts. On the sixth day, several ‘leak in IAB circuit’ alarms appeared. The nursing staff checked the tubing and refilled the balloon, as described in the manual, without notifying the medical staff or perfusionist. A few hours later, the balloon showed a leak, as blood was seen in the catheter tubing. This leak resulted in considerable damage to the device caused by a large amount of blood migrating back to the tubing and flooding the internal drive system, due to a failing ‘blood detection’ sensor. The balloon catheter was removed percutaneously and intra-aortic counterpulsation was discontinued. The patient died 20 days later of heart failure. This raises doubts over the adequacy of the protective sensors of such devices and, importantly, how to interpret the present user manuals and the instructions for troubleshooting.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3