Evaluation of a preprimed microporous hollow-fiber membrane for rapid response neonatal extracorporeal membrane oxygenation

Author:

Walczak Richard1,Lawson D Scott2,Kaemmer David2,McRobb Craig2,McDermott Patty2,Smigla Greg2,Shearer Ian2,Lodge Andrew3,Jaggers James3

Affiliation:

1. Perfusion Services, Duke University Health Systems, Durham, NC, USA,

2. Perfusion Services, Duke University Health Systems, Durham, NC, USA

3. Department of Thoracic Surgery, Duke University Health Systems, Durham, NC, USA

Abstract

Delays in initiating extracorporeal membrane oxygenation (ECMO) in the critically ill pediatric patient may lead to adverse outcomes. Maintaining a primed ECMO circuit can considerably reduce the initiation time. The predominant concerns precluding this practice are a decrease in oxygenator efficiency due to the saturation of microporous hollow fibers and compromised sterility when the oxygenator has been primed for 30 days. For institutions using a hollow-fiber oxygenator for ECMO, there are no data reporting pre-primed hollow-fiber oxygenator viability. This study reports the efficiency of oxygen transfer and the sterility of the Carmeda Minimax Plus (Medtronic, Inc, Minneapolis, MN) oxygenator after being crystalloid primed for 30 days. A total of 10 Minimax Plus oxygenators were tested for oxygen transfer in a laboratory setting utilizing fresh whole bovine blood. The control group ( n=5) were tested immediately after priming. The test group ( n=5) were oxygenators primed for 30 days with crystalloid solution and left stagnant until tested. Prior to testing, all oxygenators were circulated for 5 min and samples drawn to test for circuit sterility. Venous inlet saturations were manipulated to achieve three levels of testing: venous saturation (SvO2) of 55% for an oxygen challenge, SvO2 of 65% to comply with AAMI standards, and SvO2 of 75% to assess oxygen transfer rates and peak PaO2 achievement. Blood flow for all tests was maintained at 2 L/min with 1:1 blood to gas flow ratio and 100% FiO2. Samples were drawn pre- and postoxygenator at 1- and 6-hour time intervals to compute actual oxygen transfer values. All cultures from the test group priming solution produced no microbial growth after 30 days of stagnant prime. Average oxygen transfer values (ml/O2/min) for the control group after 1 hour of continuous use were 130.1±15.5 (@55% SvO2), 113.7±10.4 (@65% SvO2),97.7±8.9 (@75% SvO2). After 6 hours, the average transfer values increased to 134.2±13.2 (@55% SvO2), 118.76±6.6 (@65% SvO2) and 98.9±8.3 (@75% SvO2). The average oxygen transfer values after 1 hour for oxygenators primed for 30 days were 114.9±10.0 (@55% SvO2), 112.4±8.2 (@65% SvO2) and 89.6±16.0 (@75% SvO2). After 6 hours of use, the average transfer values all decreased to 111.4±2.1 (@55% SvO2, p <0.05 versus control), 104.0±5.6 (@65% SvO2, p<0.05 versus control) and 88.4±3.2 (@75% SvO2, p<0.05 versus control). In conclusion, there was a decrease in the average oxygen transfer values for the test group after 6 hours versus the control. The modest loss of oxygen transfer ability observed can be considered acceptable due to the amount of surface area of the Minimax Plus oxygenator when used on a neonate, making it feasible to adopt the practice of prepriming the Minimax oxygenator for neonatal ECMO.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3