Comparison of the effect of venovenous versus venoarterial extracorporeal membrane oxygenation on renal blood flow in newborn lambs

Author:

Ingyinn Ma1,Rais-Bahrami Khodayar2,Evangelista Rebecca3,Hogan Inger2,Rivera Oswaldo4,Mikesell Gerald T2,Short Billie L2

Affiliation:

1. Department of Neonatology,Children’s National Medical Center and The George Washington University School of Medicine, Washington, DC, USA, ,

2. Department of Neonatology,Children’s National Medical Center and The George Washington University School of Medicine, Washington, DC, USA

3. Department of Pediatric Surgery, Children’s National Medical Center and The George Washington University School of Medicine, Washington, DC, USA

4. Department of Biomedical Engineering, Children’s National Medical Center and The George Washington University School of Medicine, Washington, DC, USA

Abstract

Venovenous extracorporeal membrane oxygenation (VV ECMO) using double lumen catheters is an alternative to venoarterial (VA) ECMO and allows for total blood flow using the patient’s cardiac output in comparison to partial blood flow provided during VA ECMO. Objective: To compare the effects of VV versus VA ECMO on renal blood flow. Design: Prospective study. Setting: Research laboratory in a hospital. Subject: Newborn lambs 1-7 days of age (n=15). Interventions: In anesthetized, ventilated lambs, fe-moral artery and vein were cannulated for monitoring and renal venous blood sampling. An ultrasonic flow probe was placed on the left renal artery for continuous renal blood flow measurements. Animals were randomly assigned to control (non-ECMO), VV ECMO and VA ECMO groups. After systemic heparinization, the animals were cannulated and studied at bypass flows of 120 mL-kg/min (partial bypass) for two hours in both ECMO groups and 200 mL/kg/min (full bypass) for an additional 30 min in the VA group. Changes in blood pressure and renal flow on ECMO and during ECMO bridge unclamping were recorded continuously. Plasma renin activity (PRA) levels were sequentially sampled. Results: Systemic blood pressure was not different in VV or VA ECMO at partial bypass flow. However, systemic blood pressure increased significantly at maximal bypass flow in the VA ECMO group. There was no change in renal flow in either VV or VA ECMO groups. PRA levels did not correlate with bypass flow change. During unclamping of the ECMO bridge, blood pressure and renal flow drop significantly in the VA group, but not in the VV group. Conclusion: VV and VA ECMO at partial bypass flows had comparable effect on blood pressure, renal blood flow and PRA level in this short-term study. However, unclamping of the ECMO bridges did differentially affect blood pressure and renal blood flow between VV and VA groups. We speculate that this repeated acute change in long-run VA ECMO support may play a role in the persistent hypertension seen in some patients.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370