Normal and abnormal trans-oxygenator pressure gradients during cardiopulmonary bypass

Author:

Fisher Anthony Richard1,Baker Mya2,Buffin Mike3,Campbell Patrick4,Hansbro Stephen5,Kennington Steven6,Lilley Angela7,Whitehorne Michael8

Affiliation:

1. Heart and Lung Department, Harefield Hospital, Harefield, London, UK,

2. Department of Clinical Perfusion, Nottingham City Hospital, Nottingham, UK

3. Perfusion Department, Northern General Hospital, Sheffield, UK

4. Perfusion Department, Glasgow Royal Infirmary, Glasgow, UK

5. Department Clinical Perfusion, Leeds General Infirmary, Leeds, UK

6. Department of Clinical Perfusion, Derriford Hospital, Plymouth, UK

7. Department of Perfusion, Royal Brompton Hospital, London, UK

8. Cardiothoracic Unit, Kings College Hospital, London, UK

Abstract

A prospective study was conducted with the aims of 1) determining the normal trans-oxygenator pressure gradient characteristics for a range of oxygenators and 2) determining the characteristics, incidence and outcome of abnormally raised gradients. The trans-oxygenator pressure gradient was monitored in 3684 patients undergoing open-heart surgery in eight different hospitals. When the normal pressure gradient was measured during cardiopulmonary bypass in mmHg/L blood flow, a constant figure was obtained which was specific for each oxygenator. This gradient was abnormally raised in 16 cases (one in every 230 cases) and was raised to such an extent in three of these cases that an emergency oxygenator changeout was required (one in every 1228 cases). Among the 16 reported incidents, three different patterns of gradient changes occurred, suggesting the possibility that there were three different aetiologies. In nine of these incidents, the pressure gradient was normal immediately upon going on bypass, but rose rapidly to a plateau value, which then returned to the normal value within 40 minutes. In three cases, the pressure gradient was raised immediately upon going on bypass and then rapidly returned to the baseline. In one case, the pressure gradient was raised immediately upon going on bypass and stayed raised throughout the operation.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 31 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3