Development of an instrument to indirectly monitor arterial pCO2 during cardiopulmonary bypass

Author:

Høgetveit Jan Olav1,Kristiansen Frode2,Pedersen Thore H2

Affiliation:

1. Department of Biomedical and Clinical Engineering, Rikshospitalet, Oslo, Norway,

2. Department of Thoracic Surgery, Rikshospitalet, Oslo, Norway

Abstract

Arterial blood carbon dioxide tension (PaCO2) during cardiopulmonary bypass (CBP) is important to the conduct of perfusion with alpha-stat or pH-stat strategy. Temperature changes during CBP complicate any attempts to monitor carbon dioxide tension in the exhaust outlet of an oxygenator (PexCO2) because CO2 becomes more soluble with decreasing temperatures. Normally, this would have been the obvious and easy choice of method to indirectly measure the patient’s PaCO2. Several tests have been performed with ordinary capnographs modified to measure pCO2 at the oxygenator exhaust gas port. These tests have shown varying degrees of precision ( Br J Anaesth 1999; 82(6): 843-46; J Extra-Corpor Technol 2003; 35(3): 218-23; Br J Anaesth 2000; 84: 536; J Extra-Corpor Technol 1994; 26: 64-67). Some of the best results have been achieved by Potger et al. ( J Extra-Corpor Technol 2003; 35(3): 218-23), who found a strong correlation between the arterial temperature-corrected PexCO2 when using a standard capnograph monitoring the PaCO2 measured from a blood gas analyser (PbCO2). Our group has developed a new instrument, especially designed for oxygenator gas exhaust monitoring. The new instrument has automatic temperature correction, enabling it to show both original and corrected pCO2 values, simultaneously. Ordinary capnograph functions, such as zeroing, flow control and calibration routines, are included. The solution consists of a pCO2 sensor module, a temperature sensor, a water trap and a dedicated PC mounted on a heart-lung machine. Since the heart-lung machine was already equipped with a computer for data logging and a temperature sensor, only a box containing the pCO2 sensor module and the water trap had to be added. The PC uses a specially written program designed to collect data, make the necessary calculations and display the results on the computer screen. A temperature correction was developed based on a linear regression analysis for a data-set of 15 patients, assuming that the deviation between the measured PexCO2 from the oxygenator exhaust outlet and the PbCO2 from the blood gas analyser was linearly dependent on arterial temperature alone. Eighty-six blood gas samples were compared to the corrected PexCO2 values. The final product displayed good qualities of stability and was accurate when temperature fluctuated from 32 to 388C, even during rewarming, which has been reported to be a problem for other PexCO2 investigations ( J Extra-Corpor Technol 2003; 35(3): 218-23).

Publisher

SAGE Publications

Subject

Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology Nuclear Medicine and imaging,General Medicine

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3