Abstract
This paper describes the design, fabrication, and validation of a paper-based diagnostic device for the rapid diagnosis of hypocalcemia in dairy cattle at the point-of-care (POC). The device incorporates a 3D printed calcium ion-selective membrane (ISM) as the sensing element for free—unbound—calcium in real bovine whole blood samples. With a linear response range of 100 mM to 97.7 μM, the sensor covers the clinically relevant concentrations of Ca2+ associated with both healthy cattle as well as those suffering from hypocalcemia. The components of the Ca2+ ion-selective electrodes were successfully translated to a paper-based device to provide a sensing platform that is simple to use, disposable, and low-cost, and is therefore well-situated for applications at the POC. The paper-based calcium sensor showed a Nernstian response between 10 mM and 100 μM and required only 12 μl of sample to perform a measurement, which can be accomplished in less than two minutes without the need for time-consuming separation steps. The performance of the paper-based Ca2+ sensor was validated using the commercially available epoc® Blood Analysis System, which provided results within 5% of the data obtained with 3D printed Ca2+-ISM integrated paper-based device.
Funder
Washington State University, College of Veterinary Medicine Research - Animal Health, Research Center - Safe Food Initiative
Publisher
The Electrochemical Society
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献