Abstract
In this study, for the first time, histamine-functionalized magnetic nanoparticles (HIS-MNP) were used as sorbent phase combined with bar adsorptive microextraction (BAµE) for the determination of methylparaben (MP), ethylparaben (EP), propylparaben (PP), and butylparaben (BP) in water samples by high performance liquid chromatography coupled with diode array detection (HPLC-DAD). Extraction and desorption conditions, including desorption time and solvent, ionic strength effect and extraction time, were investigated. The extractions were performed for 90 min, 30% of NaCl and sample pH 5. Desorption was carried out using 250 μL of acetonitrile:methanol (25:75 v/v) for 30 min. The limits of detection (LODs) were of 1.5 μg L−1 and limits of quantification (LOQs) were 5.0 μg L−1 for all analytes. The intra-day precision ranged from 1% for EP, PP and BP to 3% for MP, EP, PP and BP. The inter-day precision ranged from 3% for EP to 18% for BP. The relative recoveries varied from 81 to 125%, and relative standard deviations (RSD) ranged between 1 and 16%.