Author:
Stragier André,Lopot František,Švára František,Polakovič Vladimír
Abstract
Dialysate conductivity is routinely used as a surrogate for dialysate sodium concentration. However, dialysis machine manufacturers apply different conductivity temperature correction coefficients. With the same conductivity in dialysis machines manufactured by different manufacturers, dialysate sodium may significantly differ. Also, electrolyte prescriptions are individualized (K, Ca, HCO3) and this is associated with another variation in dialysate sodium in the order of 1–5 mmol/L and both deviations are cumulative and chronic for each patient. Equivalence of the prescribed dialysate sodium and the concentration measured in it is not granted. Both variables differ and it is machine dependent! This paper analyses those variations from a technical point of view and suggests how to detect them and how to deal with or avoid them in clinical practice.
Subject
Nephrology,Hematology,General Medicine
Cited by
20 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献