The Science of Extracorporeal Ultrafiltration: Introducing a Novel Miniaturized Device

Author:

Sgarabotto Luca,Kazory Amir,Brendolan Alessandra,Di Lullo Luca,Zanella Monica,Ronco Claudio

Abstract

Introduction: Fluid overload has been associated with untoward outcomes in a variety of clinical settings. Isolated extracorporeal ultrafiltration (UF) allows for mechanical extraction of excess fluid and optimization of volume status without the established risks associated with use of high-dose diuretics. Conventional machines for renal replacement therapy can be used to perform isolated UF. However, they typically need high blood flow rates with high circuit volumes and the therapy has to be performed by trained nurses. Herein, we describe a novel device, the Artificial Diuresis-1, or AD 1 (Medica S.p.A., Medolla, Italy), which is a portable technology designed to perform extracorporeal UF at bedside. Materials and Methods: The AD 1 uses a polysulfone mini-filter to generate ultrafiltrate with the help of two forces: blood flow (Qb) and gravity (based on the height at which the ultrafiltrate collection bag is placed). In vitro experiments were performed using human blood to evaluate vascular access pressures and ultrafiltrate volumes using various central venous catheters (CVCs; 12 Fr bilume, 10 Fr with 2 separate lumens, pediatric catheter 7 Fr). A variety of combinations were tested with Qb of 20, 35, 50 mL/min and collection bag height at 20, 40, 60 cm, measuring the UF rate per minute while monitoring the pressures in the venous and arterial lines and filtration fraction. Results: The device’s performance was as expected. Regarding the pediatric CVC, it was possible to perform measurements only with a Qb of 20 mL/min due to increased venous pressure. UF rates when lines were directly connected to the blood container as well as for CVC Tesio ranged from 3.7 to 11 mL/min, for the CVC Niagara™ from 4.5 to 12.5 mL/min, and for the CVC 7 Fr from 8.5 to 10 mL/min. The pressures of the vascular accesses were kept within a range of −5/-40 mm Hg for the artery and +10/+70 mm Hg for the vein. The highest venous pressure values were found with the CVC 7 Fr (+80/+100 mm Hg). Conclusions: This novel device allows to treat patients with fluid overload in a variety of settings, from low-intensity department such as long-term care facilities to the intensive care unit. The device is small and portable, has a simple design, and is user friendly. Future studies will be needed to evaluate whether gentle UF and treatment of volume overload will translate into improvement in clinical outcomes such as a reduction in congestion-related hospital admissions.

Publisher

S. Karger AG

Subject

Urology,Cardiology and Cardiovascular Medicine

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3