Author:
Lin Shawn R.,Miller Kevin M.
Abstract
Purpose: To discuss problems associated with the implantation of two Morcher iris diaphragm models. Methods: We describe the history, intraoperative complications, and postoperative complications of 5 patients with specific Morcher iris implants. Results: We implanted Morcher 50D devices in 1 patient and Morcher 96S devices in 4 patients. Complications included postoperative rotation, device mis-sizing, difficult intraoperative rotation, zonular dehiscence, and intraoperative hemorrhage. Conclusion: Artificial iris implantation has a steep learning curve. With widespread availability on the horizon in the United States, the sharing of surgical experiences is key to achieving the best outcomes for patients.
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献