Design of Manifold with Pressure Controller for Automatic Exchange of Oxygen Gas Cylinders in Hospital

Author:

Hendryani Atika1ORCID,Nurdinawati Vita1,Dharma Nashrul1

Affiliation:

1. Jurusan Teknik Elektromedik, Politeknik Kesehatan Kementerian Kesehatan Jakarta II,, Indonesia

Abstract

The regulation and supply of oxygen as one of the medical gases in the hospital is important to ensure the availability of these gases for the survival of patients. The regulation of oxygen gas in hospitals usually uses a piping system with manifolds. The manifold will monitor the oxygen gas pressure on each tube. Manifold systems that are widely used in general can only monitor pressure but cannot perform an automatic exchange on gas cylinders if the pressure is under the permissible conditions. The manifold system design developed is equipped with pressure monitoring for automatic exchange of oxygen gas cylinders using pressure sensors and microprocessors.  The test results of the system using regulator and barometer comparisons showed the percentage value of sensor pressure accuracy of 96.92 percent and 97.16 percent. At pressure below the limit of 285 KPa manifold can perform the exchange of active gas cylinders automatically. These results show the manifold design built can work quite well.

Publisher

Institute of Research and Community Services Diponegoro University (LPPM UNDIP)

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3