Aim.The main purpose of this study was to develop a simple, precise, rapid and accurate method for the quantification of metoprolol and meldonium in human plasma. Materials and methods. The resolution of peaks of metoprolol was best achieved with Discovery C18, 50 × 2.1 mm, 5 μm column and meldonium - ZORBAX HILIC Plus, 50 × 2.1 mm, 3.5 μm column. Samples of metoprolol were chromatographed in a gradient mode (eluent A (acetonitrile – water – formic acid, 5 : 95 : 0.1 v/v), eluent B (acetonitrile – formic acid, 100 : 0.1 v/v)). The initial content of the eluent B is 0%, which increases linearly by 1.0 min to 100% and to 1.11 min returns to the initial 0%. The mobile phase was delivered at a flow rate of 0.400 mL/min into the mass spectrometer ESI chamber. The injection volume was 5μl. Samples of meldonium were chromatographed in a isocratic using mobile phase water – acetonitrile – ammonium formate buffer 200 мМ, 20 : 75 : 5 v/v). Results.The total chromatographic run time was 2.0 minutes and the elution of metoprolol, meldonium and IS occurred at ~1.39 and 1.18 minutes, respectively.A linear response function was established at 2 - 200 ng/mL for metoprolol and 50 -5000 ng/mL for meldonium in human plasma. The% mean recovery for metoprolol in LQC, MQC and HQC was 99.0%, 107.5% and 96.8%, for meldonium in LQC, MQC and HQC was 94.1%, 100.2% and 93.1% respectively. The lowest concentration with the RSD <20% was taken as LLOQ and was found to be 2.31 ng/mL for metoprolol, 47.70 ng/mL for meldonium. The % accuracy of LLOQ samples prepared with the different biological matrix lots were found 115.4% for metoprolol and 95.5% for meldonium, which were found within the range of 80.00–120.00% for the seven different plasma lots. % CV for LLOQ samples was observed as 12.8% and 7.7% respectively, which are within 20.00% of the acceptance criteria. Conclusion.A rapid method was developed for simultaneous determination of metoprolol and meldonium in human plasma. The method was strictly validated according to the ICH guidelines. Acquired results demonstrate that proposed strategy can be effortlessly and advantageously applied for routine examination of metoprolol and meldonium in human plasma.

Publisher

Pensoft Publishers

Subject

Pharmacology (medical),Pharmaceutical Science,Pharmacy

Reference26 articles.

1. HPLC quantification of metoprolol with solid-phase extraction for the drug monitoring of pediatric patients

2. Gas Chromatographic-Mass Spectrometric Differentiation of Atenolol, Metoprolol, Propranolol, and an Interfering Metabolite Product of Metoprolol

3. Development, validation and analytical error function of two chromatographic methods with fluorimetric detection for the determination of bisoprolol and metoprolol in human plasma

4. EMEA/CHMP. Guideline on Validation of Bioanalytical Methods (Draft). EMEA/CHMP/EWP/192217/2009. European Medicines Agency/Committee for Medicinal Products for Human Use, 2009. Available from: http://www.ema.europa.eu/docs/en_GB/document_library/Scientific_guideline/2009/12/WC500018062.pdf [accessed 19 November 2009]

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3