Downloadable Software Algorithm Reduces Inappropriate Shocks Caused by Implantable Cardioverter-Defibrillator Lead Fractures

Author:

Swerdlow Charles D.1,Gunderson Bruce D.1,Ousdigian Kevin T.1,Abeyratne Athula1,Sachanandani Haresh1,Ellenbogen Kenneth A.1

Affiliation:

1. From the Department of Cardiology (C.D.S.), Cedars-Sinai Heart Institute, Cedars-Sinai Medical Center, Los Angeles, Calif; Medtronic, Inc (B.D.G., K.T.O., A.A., H.S.), Mounds View, Minn; and Division of Cardiology (K.A.E.), Medical College of Virginia/Virginia Commonwealth University, Richmond, Va.

Abstract

Background— Downloadable software upgrades are common in consumer electronics but not in implantable medical devices. Fractures of implantable cardioverter-defibrillator (ICD) leads present commonly as inappropriate shocks. A lead-integrity alert (LIA) designed to reduce inappropriate shocks is the first software download approved to enhance nominally functioning, previously implanted ICDs. Methods and Results— We performed a prospective study to determine whether an LIA could reduce inappropriate shocks. Patients were included if they had ICD lead fractures confirmed by analysis of explanted leads. The LIA group included the first 213 patients who met the inclusion criteria after the LIA was approved who had the LIA downloaded. The LIA is triggered either by high impedance or rapid oversensing. It responds by delaying detection of ventricular fibrillation and initiating a patient alert every 4 hours. The control group included the first 213 patients who did not have the LIA downloaded. They were monitored by conventional daily impedance measurements that respond with a daily alert. The LIA group had a 46% relative reduction (95% confidence interval 34% to 55%) in the percentage of patients with ≥1 inappropriate shock (LIA 38% versus control 70%, P <0.001) and a 50% relative reduction (95% confidence interval 33% to 61%) in the percentage with ≥5 shocks (25% versus 50%, P <0.001). The LIA group also had a higher percentage of patients who either did not receive a shock or had ≥3 days of warning before the shock (72% versus 50%, P <0.001). Conclusions— A software download that upgrades previously implanted ICDs without surgical revision reduces inappropriate shocks caused by lead fractures.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3