Stent Thrombogenicity Early in High-Risk Interventional Settings Is Driven by Stent Design and Deployment and Protected by Polymer-Drug Coatings

Author:

Kolandaivelu Kumaran1,Swaminathan Rajesh1,Gibson William J.1,Kolachalama Vijaya B.1,Nguyen-Ehrenreich Kim-Lien1,Giddings Virginia L.1,Coleman Leslie1,Wong Gee K.1,Edelman Elazer R.1

Affiliation:

1. From the Harvard–MIT Division of Health Sciences and Technology, Cambridge, MA (K.K., R.S., W.J.G., V.B.K., G.K.W., E.R.E.); Cardiovascular Division, Brigham and Women's Hospital, Harvard Medical School, Boston, MA (K.K., E.R.E.); Abbott Vascular, Santa Clara, CA (K.-L.N.-E., V.L.G., L.C.); and Concord Biomedical Sciences and Emerging Technologies, Lexington, MA (G.K.W.).

Abstract

Background— Stent thrombosis is a lethal complication of endovascular intervention. Concern has been raised about the inherent risk associated with specific stent designs and drug-eluting coatings, yet clinical and animal support is equivocal. Methods and Results— We examined whether drug-eluting coatings are inherently thrombogenic and if the response to these materials was determined to a greater degree by stent design and deployment with custom-built stents. Drug/polymer coatings uniformly reduce rather than increase thrombogenicity relative to matched bare metal counterparts (0.65-fold; P =0.011). Thick-strutted (162 μm) stents were 1.5-fold more thrombogenic than otherwise identical thin-strutted (81 μm) devices in ex vivo flow loops ( P <0.001), commensurate with 1.6-fold greater thrombus coverage 3 days after implantation in porcine coronary arteries ( P =0.004). When bare metal stents were deployed in malapposed or overlapping configurations, thrombogenicity increased compared with apposed, length-matched controls (1.58-fold, P =0.001; and 2.32-fold, P <0.001). The thrombogenicity of polymer-coated stents with thin struts was lowest in all configurations and remained insensitive to incomplete deployment. Computational modeling–based predictions of stent-induced flow derangements correlated with spatial distribution of formed clots. Conclusions— Contrary to popular perception, drug/polymer coatings do not inherently increase acute stent clotting; they reduce thrombosis. However, strut dimensions and positioning relative to the vessel wall are critical factors in modulating stent thrombogenicity. Optimal stent geometries and surfaces, as demonstrated with thin stent struts, help reduce the potential for thrombosis despite complex stent configurations and variability in deployment.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 686 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3