In Vitro Identification of Angioplasty-Induced Injury by Use of Vascular Acoustic Emissions

Author:

Vonesh Michael J.1,Mockros Lyle F.1,Davidson Charles J.1,Chandran Krishnan B.1,McPherson David D.1

Affiliation:

1. the Department of Medicine, Section of Cardiology (M.J.V., C.J.D., D.D.M.), and the Department of Biomedical Engineering (M.J.V., L.F.M.), Feinberg Cardiovascular Research Institute, Northwestern University, Chicago, Ill, and the College of Engineering, Department of Biomedical Engineering (K.B.C.), University of Iowa, Iowa City.

Abstract

Background We have developed a novel method of diagnosing stress-induced vascular injury. This approach uses the sound energy released from atherosclerotic arterial tissue during in vitro balloon angioplasty to characterize type and severity of induced trauma. Methods and Results Thirty-two postmortem human peripheral arterial specimens 1.0 cm long were subjected to in vitro balloon angioplasty with simultaneous acoustic emission monitoring. Specimens were examined before and after angioplasty to ascertain the extent of angioplasty-induced injury. Gross observation was used to identify dissection. A three-dimensional intravascular ultrasound reconstruction technique was used to estimate the luminal surface area of the specimen. Change in luminal surface area (postangioplasty minus preangioplasty) was used to quantify induced injury. The energy content and spectral distribution of the digitally acquired vascular acoustic emission (VAE) signals were computed. Comparisons of angioplasty-induced trauma with VAE signal characteristics were made. Dissection (mural laceration of variable depth) was observed in 15 of 32 specimens. Eleven showed no evidence of induced dissection, and 6 had preexisting intimal disruptions. The energy content of the VAE signals collected from specimens with dissection was greater than that obtained from those in which dissection was absent: 845±89.4 mJ (mean±SEM; n=15) versus 128±40.8 mJ (n=11; P <.001). Comparison of induced trauma and VAE signal energy demonstrated a proportional relationship ( r =.87, P <.001, n=32). Conclusions VAE signals contain information characterizing type and severity of angioplasty-induced arterial injury. Because vascular injury is related to adverse procedural outcome, development of VAE technology as an adjunct to conventional diagnostic modalities may facilitate optimal balloon angioplasty delivery and postprocedural care.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3