Improved Defibrillation Threshold With a New Epicardial Carbon Electrode Compared With a Standard Epicardial Titanium Patch

Author:

Alt Eckhard U.1,Fotuhi Parwis C.1,Callihan Richard L.1,Rollins Dennis L.1,Mestre Edgar1,Combs Marty P.1,Smith William M.1,Ideker Raymond E.1

Affiliation:

1. From the Departments of Medicine (R.L.C., W.M.S., R.E.I.) and Pathology (D.L.R., R.E.I.), Duke University Medical Center, Durham, NC; the Engineering Research Center for Emerging Cardiovascular Technologies, the Department of Biomedical Engineering of the School of Engineering (W.M.S., R.E.I), Duke University, Durham, NC; and the I. Medical Clinic (E.U.A., E.M., M.P.C), Technische Universität Munich, Germany.

Abstract

Background Recent studies show that depending on the type of shock morphology used, 5% to 15% of patients requiring implantable defibrillators cannot be treated with a nonthoracotomy system. In these cases, an epicardial patch–based system becomes necessary. In this study, we investigated a newly developed epicardial carbon electrode as an alternative to a standard epicardial titanium patch. Methods and Results A tubular epicardial braided carbon electrode of 7F diameter and 14-cm length applied in a U-shape to the epicardium was compared with a standard left ventricular epicardial 15-cm 2 titanium mesh patch (CPI Inc). As cathode, a CPI endocardial lead, a Medtronic lead, or a carbon-platinum-iridium prototype electrode was used. Ventricular fibrillation was induced with a 60-Hz generator and allowed to continue for 10 seconds before a shock was given. Two different biphasic shock waveforms (3.2/2- and 6/6-millisecond) were delivered by the six electrode configurations. Eight dogs (weight, 24.5±1.3 kg) underwent an up-down defibrillation protocol. The order of testing the epicardial electrodes, the endocardial cathodes, and the waveform was randomized. With the epicardial carbon electrode, the mean defibrillation threshold (DFT) energy decreased 39% to 56% and the voltage decreased 24% to 35% compared with the titanium patch: from 8.3±2.5 to 4.9±3.6 J with the CPI lead and the 3.2/2-millisecond waveform, from 6.2±2.5 to 2.9±2.1 J with the carbon-platinum-iridium prototype, and from 6.4±3.4 J to 3.5±2.6 J with the Medtronic lead ( P ≤.05). The DFT determinations with the 6/6-millisecond biphasic waveform showed a similar trend with slightly higher values. Conclusions Compared with a titanium patch, the new braided epicardial electrode significantly decreases the defibrillation energy requirements. This effect can be maximized by using an endocardial carbon-platinum-iridium prototype as cathode and a short duration biphasic waveform.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术