Accuracy of a continuous blood pressure monitor based on arterial tonometry.

Author:

Sato T1,Nishinaga M1,Kawamoto A1,Ozawa T1,Takatsuji H1

Affiliation:

1. Department of Physiology, Kochi Medical School, Japan.

Abstract

A validation study of the continuous noninvasive tonometric blood pressure monitor called JENTOW was performed in 20 normotensive subjects and 10 hypertensive patients. Tonometric and intra-arterial blood pressures were simultaneously recorded at supine rest and during a Valsalva maneuver and tilting test. The results of the strict evaluation of the instrument's capacity for reproducing intra-arterial blood pressure were as follows: 1) The overall frequency response of the transcutaneous blood pressure-monitoring system based on arterial tonometry was flat, with negligible delay to intra-arterial blood pressure in the range of 0-5 Hz. 2) The largest discrepancy between intra-arterial and tonometric pressure waveforms was found at the early systolic phase; except for this phase, the tonometric waveform was almost equal to the intra-arterial waveform. 3) The beat-to-beat variability of tonometric pressure corresponded to that of intra-arterial pressure almost perfectly in the physiologically significant frequency range of 0-0.5 Hz. 4) During resting conditions, the averages of the systolic and diastolic values measured tonometrically corresponded well to those measured intra-arterially. 5) The changes in the between-method discrepancy of blood pressure values during the Valsalva maneuver were statistically significant but small (< 5 mm Hg). 6) No significant effect of postural tilting was found on the between-method discrepancy. We conclude that this method is clinically acceptable and reliable except for its limited capacity for recording the higher frequency intra-arterial waveform and for responding to the relatively rapid and large transient changes in blood pressure.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Internal Medicine

Reference13 articles.

1. A transducer for the external measurement of arterial blood pressure;Pressman GL;IEEE Trans Biomed Electron,1963

2. Drzewiecld GM Melbin J Noordergraaf A: Arterial tonometry: Review and analysis. / Biomech 1983;16:141-152

3. Eckerle JS: Tonometry arterial in Webster JG (ed): Encyclopedia of Medical Devica and Instrumentation. New York John Wiley & Sons Inc 1988 pp 2770-2776

4. Bcndat JS Piersol AG: Random Data: Analysis and Measurement Procedures. New York John Wiley & Sons Inc 1971