Affiliation:
1. University of Iowa Division of Cardiology, Department of Medicine, Iowa City, IA
Abstract
Background
Venous stenosis is a common complication of transvenous lead implantation, but the risk factors for venous stenosis have not been well defined to date. This study was designed to evaluate the incidence of and risk factors for venous stenosis in a large consecutive cohort.
Methods and Results
A total of 212 consecutive patients (136 male, 76 female; mean age 69 years) with existing pacing or implantable cardioverter‐defibrillator systems presented for generator replacement, lead revision, or device upgrade with a mean time since implantation of 6.2 years. Venograms were performed and percentage of stenosis was determined. Variables studied included age, sex, number of leads, lead diameter, implant duration, insulation material, side of implant, and anticoagulant use. Overall, 56 of 212 patients had total occlusion of the subclavian or innominate vein (26%). There was a significant association between the number of leads implanted and percentage of venous stenosis (
P
=0.012). Lead diameter, as an independent variable, was not a risk factor; however, greater sum of the lead diameters implanted was a predictor of subsequent venous stenosis (
P
=0.009). Multiple lead implant procedures may be associated with venous stenosis (
P
=0.057). No other variables approached statistical significance.
Conclusions
A significant association exists between venous stenosis and the number of implanted leads and also the sum of the lead diameters. When combined with multiple implant procedures, the incidence of venous stenosis is increased.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Cardiology and Cardiovascular Medicine
Cited by
70 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献