Affiliation:
1. the University Department of Medicine, City Hospital, Birmingham, England.
Abstract
Background and Purpose
Ambulatory blood pressure monitoring (ABPM) devices are increasingly used in the assessment of hypertension, but their value in patients after a stroke is unknown, despite the fact that hypertension is an important cause of stroke and many patients have relatively high blood pressure (BP) levels at presentation. We therefore investigated the clinical use of a 24-hour oscillometric ABPM device in patients after acute stroke. We also investigated ABPM in different types of stroke (thrombosis, hemorrhage, and transient ischemic attack) and ethnic and sex differences.
Methods
BP was measured manually with a standard mercury sphygmomanometer, and ABPM measurements were made with an oscillometric device. The first reading obtained with the ABPM device was compared with simultaneous manual BP measurements. Mean daytime and nighttime pressures were also analyzed to determine the frequency of nocturnal BP falls (“dipping”).
Results
We studied 86 patients (48 men; mean±SD age, 64.2±9.2 years) admitted with acute-onset stroke (ictus within 12 hours) in a district general hospital. Thirty-one patients (36.0%) had a previous history of hypertension. The median percentage of successful BP readings by ABPM was 92% (interquartile range, 72 to 98). There was no significant difference in manual BP levels compared with the first simultaneous systolic or diastolic ABPM measurements. Systolic BPs recorded by ABPM were significantly higher in black patients with acute stroke and in patients with intracerebral hemorrhage, who also showed a trend toward higher nocturnal BPs. There was no difference in BPs between men and women and those who were alive or dead 6 months later (
P
=NS). There was also no difference between mean day and night systolic BP (mean difference, 1.9 mm Hg;
P
=.08), although mean daytime diastolic BP was higher than mean nighttime diastolic BP (mean difference, 2.4 mm Hg;
P
=.01). Patients with stroke therefore demonstrated a loss of diurnal BP rhythm and may be considered “nondippers”; there was also a trend toward “reverse dipping” in patients with intracranial hemorrhage.
Conclusions
This study demonstrates higher systolic BPs as recorded by ABPM (but not manually) in patients with intracerebral hemorrhage than in those with cerebral infarcts; higher levels were also found in blacks. ABPM recordings are useful in the assessment of BP in patients with stroke, who may be considered nondippers.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Neurology (clinical)
Reference28 articles.
1. Blood Pressure After Stroke
2. Coats AJS Conway J Sleight P. Ambulatory monitoring: clinical trial size and precision. J Hypertens. 1989;7(suppl 6):s357. Abstract.
3. Coats AJS. Reproducibility or variability of casual and ambulatory blood pressure data: implications for clinical trials. J Hypertens. 1990;6(suppl 6):s17-s20.
4. O'Brien E Mee F Atkins N O'Malley K. Accuracy of the Spacelabs 90207 determined by the British Hypertension Society protocol. J Hypertens. 1991;9(suppl 5):s25-s31.
Cited by
62 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献