Abstract
Background and Purpose
The use of left ventricular assist devices has become an established method in bridging patients with end-stage cardiac failure to heart transplantation. Since thromboembolism is one of the major complications, we undertook this study to evaluate the clinical significance of Doppler microembolic signals (MES) in patients with left ventricular assist devices.
Methods
Six patients with left ventricular assist devices were monitored for MES with transcranial Doppler ultrasonography during the first 30 postoperative days. Additionally, repeated (10 per day and patient) and prolonged (3 hours per patient) monitorings were performed to assess the adequacy of the 30-minute recordings. Three observers evaluated 30 randomly assigned monitorings in a blinded fashion to assess the interobserver variability. The relation between MES counts and clinical, radiological, hemostaseological, and pump flow parameters and the predictive value of MES counts regarding the occurrence of embolic events was evaluated.
Results
Ten ischemic cerebrovascular accidents and 2 peripheral thromboembolic events occurred during the observation period of 177 days (total incidence, 6.8%). MES were found in 143 of 170 monitorings (84.1%). Their counts were significantly higher on days with clinically manifest embolic events as compared with event-free days (18.5 [3-74] versus 4 [0-52], respectively, median and 95% CI;
P
<.001, Mann-Whitney). The predictive value of MES counts above 7 per 30 minutes was high (75%). Significant differences in the incidence and counts of MES as well as in the incidence of clinically manifest embolic events were noted among the six patients (all
P
<.01) without equal differences in anticoagulant treatment or pump flow. Interobserver agreement was high (
P
=.78 to .89, unpaired Student’s
t
test). Considerable short- and long-term intrapatient variations of MES counts, without consistent pattern, were noted.
Conclusions
Serial monitoring for MES is prognostically superior to single monitorings in patients with left ventricular assist devices. In the future, this new application mode may individually guide anticoagulation strategies and even influence the decision regarding early cardiac transplantation versus long-term use of the assist devices.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Clinical Neurology
Reference31 articles.
1. Annual report of the US Scientific Registry for organ transplantation and the organ procurement and transplantation network—1990. Es 8.-9. Washington DC: US Department of Health and Human Services.
2. Eight years' experience with bridging to cardiac transplantation
3. Ventricular Assist and Total Artificial Heart Devices for Clinical Use in 1993
4. Innovative Ventricular Assist Systems
Cited by
56 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献