Affiliation:
1. From the Heart Disease Department (F.A., F.C., M.S.) and the Neurological Sciences Department (G. di B., F.F.G.), S. Filippo Neri Hospital, Rome, Italy.
Abstract
Background and Purpose—
We sought to determine whether the introduction of EEG monitoring during head-up tilt testing could significantly improve the understanding of the cerebral events occurring during tilt-induced vasovagal syncope and the potential danger to the patient of this diagnostic procedure.
Methods—
EEG monitoring was performed during head-up tilt testing in a group of 63 consecutive patients (27 males and 36 females; mean age, 41.5 years) with a history of recurrent syncope of unknown origin despite extensive clinical and laboratory assessment.
Results—
Syncope occurred in 27 of 63 patients (42.8%) during head-up tilt testing and was found to be cardioinhibitory in 11 of 27 (40.7%) and vasodepressor in 16 of 27 (59.3%). All patients with a negative response to head-up tilt testing showed no significant EEG modifications. In patients with vasodepressor syncope, a generalized high-amplitude, 4- to 5-Hz (theta range) slowing of EEG activity appeared at the onset of syncope, followed by an increase of brain-wave amplitude with the reduction of frequency at 1.5 to 3 Hz (delta range). The return to the supine position was associated with brain-wave amplitude reduction and frequency increase to 4 to 5 Hz, followed by restoration of a normal EEG pattern and arousal (mean total duration of syncope, 23.2 seconds.). In patients with cardioinhibitory syncope, a generalized high-amplitude EEG slowing in the theta range was noted at the onset of syncope, followed by a brain-wave amplitude increase and slowing in the delta range. A sudden reduction of brain-wave amplitude then ensued, leading to the disappearance of electrocerebral activity (“flat” EEG). The return to the supine position did not allow either the immediate resolution of EEG abnormalities or consciousness recovery, both of which occurred after a further time interval (mean total duration of syncope, 41.4 seconds.).
Conclusions—
EEG monitoring during head-up tilt testing allowed recording and systematic description of electrocerebral abnormalities developing in the course of tilt-induced vasovagal syncope.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Neurology (clinical)
Cited by
51 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献