Feasibility of Redo-Transcatheter Aortic Valve Replacement in Sapien Valves Based on In Vivo Computed Tomography Assessment

Author:

Fukui Miho1ORCID,Okada Atsushi2ORCID,Thao Kiahltone R.2,Burns Marcus R.3,Koike Hideki1,Wang Cheng2ORCID,Phichaphop Asa2,Lesser John R.23,Sorajja Paul23ORCID,Cavalcante João L.13,Bapat Vinayak N.23

Affiliation:

1. Cardiovascular Imaging Research Center and Core Lab (M.F., H.K., J.L.C.), Minneapolis Heart Institute Foundation, MN.

2. Valve Science Center (A.O., K.R.T., C.W., A.P., J.R.L., P.S., V.N.B.), Minneapolis Heart Institute Foundation, MN.

3. Minneapolis Heart Institute at Abbott Northwestern Hospital, MN (M.R.B., J.R.L., P.S., J.L.C., V.N.B.).

Abstract

BACKGROUND: Our aim was to assess the feasibility of repeat transcatheter aortic valve (TAV) replacement for degenerated Sapien3 (S3) prostheses by simulating subsequent implantation of S3 or Evolut, using in vivo computed tomography–based sizing and the impact on coronary and patient-prosthesis mismatch risks. METHODS: Computed tomography scans from 356 patients with prior S3 TAV replacement implantation were analyzed. The in vivo sizing for second TAV based on averaged area of 3 levels of outflow, mid (narrowest) and inflow, was compared with in vitro recommendations, that is, same size as index S3 for second S3 and 1 size larger for Evolut. Risks of coronary obstruction and patient-prosthesis mismatch were determined by valve-to-aorta distance and estimated effective orifice area, respectively. RESULTS: Overall, the majority of patients (n=328; 92.1%) had underexpanded index S3 with an expansion area of 94% (91%–97%), leading to significant differences in size selection of the second TAV between in vivo and in vitro sizing strategies. Expansion area <89% served as a threshold, resulting in 1 size smaller than the in vitro recommendations were selected in 45 patients (13%) for S3-in-S3 and 13 (4%) for Evolut-in-S3, while the remaining patients followed in vitro recommendations ( P <0.01, in vivo versus in vitro sizing). Overall, 57% of total patients for S3-in-S3 simulation and 60% for Evolut-in-S3 were considered low risk for coronary complications. Deep index S3 implantation (odds ratio, 0.76 [interquartile range, 0.67–0.87]; P <0.001) and selecting Evolut as the second TAV (11% risk reduction in intermediate- or high-risk patients) reduced coronary risk. Estimated moderate or severe patient-prosthesis mismatch risk was 21% for S3-in-S3 and 1% for Evolut-in-S3, assuming optimal expansion of the second TAV. CONCLUSIONS: Redo-TAV replacement with S3-in-S3 and Evolut-in-S3 could be feasible with low risk to coronaries in ≈60% of patients, while the remaining 40% will be at intermediate or high risk. The feasibility of redo-TAV replacement is influenced by sizing strategy, type of second TAV, native annular anatomy, and implant depth.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3