Treatment of Acute Stanford Type B Aortic Dissection With a Novel Cylindrical Balloon Catheter in Dogs

Author:

Terai Hiromu1,Tamura Nobushige1,Nakamura Tatsuo1,Nishimura Kazunobu1,Tsutsui Norimasa1,Shimizu Yasuhiko1,Komeda Masashi1

Affiliation:

1. From the Department of Cardiovascular Surgery, Graduate School of Medicine, Kyoto University (H.T., N. Tamura, K.N., M.K.); Institute for Frontier Medical Sciences, Kyoto University (T.N., Y.S.); and Tokai Medical Co (N. Tsutsui).

Abstract

Background —Despite recent progress in medical and surgical treatment, acute type B aortic dissection still carries a high mortality rate. We have developed a novel cylindrical balloon catheter for less invasive treatment to block the entry of the dissection and induce thrombotic occlusion of the false lumen. The balloon has the shape of a sheet when deflated but a double-cylinder shape when inflated. Therefore, aortic blood flow is maintained through the cylindrical lumen during balloon inflation. Methods and Results —Six beagle dogs underwent a left thoracotomy at the 6th intercostal space. An acute dissection of 4-cm length was created surgically on the descending aorta. The balloon catheter was inserted through the distal descending aorta and advanced to the entry site. The balloon catheter was inflated for 6 hours. The blood flow in the descending aorta and the position of the balloon was monitored by color Doppler echovasculography. Four dogs were killed humanely on the following day and 2 dogs 10 days after the surgery. The descending aorta was examined macroscopically and microscopically in all dogs. In all dogs, the false lumen was occluded by thrombi. Although no dog had clinical evidence of distal thromboembolism, 2 of the 4 dogs that were killed on the second postoperative day had fresh mural thrombi in the true lumen. Conclusions —The false lumen of the acute type B aortic dissection was effectively occluded by the novel cylindrical balloon catheter in the canine experimental model. The thrombus formation in the true lumen is the problem to be solved.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Reference23 articles.

1. Glower DD Fann JI Speier RH et al. Comparison of medical and surgical therapy for uncomplicated descending aortic dissection. Circulation . 1990;82(suppl IV):IV-39–IV-46.

2. Svensson LG Crawford ES Hess KR et al. Dissection of the aorta and dissecting aortic aneurysms: improving early and long-term surgical results. Circulation . 1990;82(suppl IV):IV-24–IV-38.

3. Operation for acute and chronic aortic dissection: recent outcome with regard to neurologic deficit and early death

4. Early and late outcome of operated and non-operated acute dissection of the descending aorta

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献