Experimental atrial septal defect closure with a new, transcatheter, self-centering device.

Author:

Das G S1,Voss G1,Jarvis G1,Wyche K1,Gunther R1,Wilson R F1

Affiliation:

1. Department of Medicine, University of Minnesota Medical School, Minneapolis 55455.

Abstract

BACKGROUND Despite two decades of research, a transcatheter atrial septal defect closure device is not available for clinical use. We have designed a new superelastic Nitinol-Dacron, double-disk, self-centering, atrial septal defect closure device and studied its efficacy in a canine model of atrial septal defects. METHODS AND RESULTS Atrial septal defects were created surgically in 20 adult dogs using either a 7.5-mm or 10-mm punch. Percutaneous transcatheter closures were attempted using a new device. The device sizes used were 20 mm in 6 dogs, 22 mm in 9, and 25 mm in 5 (22.1 +/- 1.9 mm, mean +/- SD). The stretched atrial septal defect diameter was 10.5 +/- 1.3 mm, and the device to stretched atrial septal defect diameter ratio was 2.1 +/- 0.3. Closures were successful in 19 studies and unsuccessful in 1. Angiography showed a left-to-right shunt in all 20 dogs before closure. Immediately after closure (n = 19), there were no shunts in 17 and trivial shunts in 2. Six dogs were followed for a period of 4.7 +/- 3.0 months (range, 2 to 8 months). The trivial shunt present in 1 animal immediately after closure had closed by the time of the repeat study. Spontaneous embolization of the device was not seen during follow-up. A solitary wire fracture was found 8 months after closure in 1 device. Light microscopy at 8 weeks in 3 dogs showed the devices to be covered by smooth endocardium, enmeshed in mature collagen tissue, with a minimal mononuclear cell infiltration. Retrievability was assessed by deliberately embolizing 4 devices in 2 dogs into the right atrium (n = 1) and pulmonary artery (n = 3). All devices were successfully retrieved with a snare. CONCLUSIONS This feasibility study demonstrates that this new self-centering atrial septal defect closure device has a number of design features that permit effective and safe closures in a canine model. These results support the investigation of this device in human clinical trials.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 170 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3